Radiation Protection Lead Glass Resistant To Most Chemicals with 55-65% Lead Oxide Content and 4mmPb Lead Equivalent for X Ray Shielding

Brand Name Jovi
Certification CE
Model Number 1-cf-15
Document Lead glass specification.pdf
Minimum Order Quantity 1 Pc
Price $2130/pc-$6380/pc
Packaging Details In Wooden Case Or Customized
Delivery Time 5-8 Work Days
Payment Terms T/T
Supply Ability 2000 Pcs Per Month
Product Details
Lead Oxide Content 55-65% By Weight Radiographic Quality ≥99.97%
Light Transmission 84-86% Impact Resistance Steel Ball Drop From 1.5 M - No Fracture
Lead Equivalent 4mmPb Minimum Feature Resolution 0.5 Mm At 30 Cm
Leave a Message
Product Description
Application Deep Dive: SPECT, PET‑CT, Hot Lab, and Therapy Isolation
SPECT (Single Photon Emission Computed Tomography)
Isotope
⁹⁹ᵐTc (140 keV, 6 hour half‑life) – accounts for >80% of SPECT studies.
Shielding requirement
2.5–3.0 mm Pb equivalent.
Jovvi solution
Monolithic lead glass, 14–18 mm physical thickness. Light transmission 80–84%.

Special consideration: SPECT cameras rotate around the patient. The control booth window must provide a wide field of view. Jovvi offers panels up to 1,500×1,200 mm – large enough to see the entire gantry.

Installation note: Wall shielding must match glass Pb eq. If walls are 2.5 mm Pb and glass is 3.0 mm Pb, the glass is not the weak point – but the wall‑to‑glass junction must be sealed with lead tape.

PET‑CT (Positron Emission Tomography + CT)
Isotope
¹⁸F‑FDG (511 keV annihilation photons, 110 minute half‑life).
Shielding requirement
4.0–5.0 mm Pb equivalent minimum, but monolithic glass at this thickness is heavy and still transmits ~50%.
Jovvi solution
Composite window – 2.5 mm Pb glass + 10 mm lead acrylic + 2.5 mm Pb glass. Total thickness ~35 mm. 511 keV transmission <15%. Weight 70 kg/m² (vs. 110 kg/m² for 5 mm monolithic).

Special consideration: PET‑CT technologists view fusion images (PET overlaid on CT) through the glass. Optical distortion must be minimal. Jovvi composite windows are optically bonded with UV‑stable adhesive – no haze or double images.

Hot Lab (Radiopharmacy)
Isotopes handled
⁹⁹ᵐTc, ¹³¹I, ¹⁸F‑FDG, ⁶⁸Ga, ¹⁷⁷Lu (varies).
Shielding requirement
2.5 mm Pb equivalent – sufficient because handling time is short (minutes per dose) and syringe shields provide additional protection.
Jovvi solution
Small panel (400×300 mm or 500×400 mm), 2.5 mm Pb, anti‑reflection coating on both sides to eliminate glare from laminar flow hood lights.

Special consideration: Pharmacists need to read 0.01 mL graduations on 1 mL syringes. Jovvi guarantees 0.5 mm minimum feature resolution at 30 cm viewing distance.

High‑Dose Therapy Isolation Rooms
Isotopes
¹³¹I (364 keV, 8 day half‑life) – typical administered activity 100–800 mCi. Also ¹⁷⁷Lu (208 keV), ⁹⁰Y (pure beta – no gamma, but bremsstrahlung).
Shielding requirement
3.5–4.0 mm Pb equivalent.
Jovvi solution
Monolithic lead glass, 20–24 mm physical thickness. Nurse station window sized 800×600 mm.

Special consideration: Patients may stay in isolation for several days. Nurses observe for distress, bleeding, or falls. Glass must be distortion‑free to allow accurate assessment of patient color and consciousness. Jovvi provides optical flatness certificate with every therapy grade panel.